How it works
Sample molecules are ionised (for peptides, usually by electrospray or MALDI), and the instrument measures the mass-to-charge ratio of the resulting ions. The compound's molecular mass is derived from the spectrum — precisely enough to distinguish molecules differing by a single amino-acid residue or less.
Why it is the identity method
Every peptide sequence has a calculable molecular mass. If the measured mass matches the calculated one, the material is consistent with the declared structure; if not, the vial contains a different molecule or a significant modification. This is why, in the "HPLC + MS" pair, chromatography covers purity while mass spectrometry covers identity.
The method's limits
A mass match is necessary but not absolute: molecules with identical mass (for example, with rearranged residues) are indistinguishable in a basic spectrum; fragmentation methods (MS/MS) exist for that. A mass spectrum also says nothing about quantity or sterility. Each property is confirmed by its own method.
References
- Electrospray Mass Spectrometry and Tandem Mass Spectrometry of Synthetic Multicomponent Peptide Mixtures: Determination of Composition and Purity — Analytical Biochemistry, 219(2), 261–277 (PMID 8080083; DOI 10.1006/abio.1994.1266), 1994.